Monte Carlo investigation of dose distribution of uniformly and non-uniformly loaded standard and notched eye plaques.


ABSTRACT: To investigate the effect of using non-uniform loading and notched plaques on dose distribution for eye plaques. Using EGSnrc Monte Carlo (MC) simulations, we investigate eye plaque dose distributions in water and in an anatomically representative eye phantom. Simulations were performed in accordance with TG43 formalism and compared against full MC simulations which account for inter-seed and inhomogeneity effects. For standard plaque configurations, uniformly and non-uniformly loaded plaque dose distributions in water showed virtually no difference between each other. For standard plaque, the MC calculated dose distribution in planes parallel to the plaque is narrower than the TG43 calculation due to attenuation at the periphery of the plaque by the modulay. MC calculated the dose behind