ABSTRACT: Background
Defibrillation threshold (DFT) testing is done to assess whether proper sensing of ventricular fibrillation and adequate safety margin for defibrillation are present in an implantable cardioverter defibrillator (ICD). This case report presents an intuitive method for lowering the DFT. It may be used on a larger scale in other patients with high DFTs when other methods for lowering the DFT (changing medications, adjusting the device, and adding coils) are not feasible or preferable to use.Case summary
A 64-year-old male presented to the emergency room with failed appropriate shocks from his ICD. Device interrogation revealed that he failed his first maximum output shock before subsequent shock at the same polarity and output succeeded, suggesting a high DFT. There