Evaluation of MR-safe bioptomes for MR-guided endomyocardial biopsy in minipigs: a potential radiation-free clinical approach.


ABSTRACT:

Background

Diagnostic accuracy of endomyocardial biopsy could improve if clinically safe magnetic resonance (MR)-compatible bioptomes were available. We explored two novel MR-compatible cardiac bioptomes for performance, safety, and clinical viability, employing in vivo minipig trials and phase-contrast synchrotron radiation computed microtomography (SRµCT).

Methods

Analysis of ex vivo obtained pig endomyocardial biopsies was performed using phase-contrast SRµCT and conventional two-dimensional histology. The technical performance was evaluated by measuring volume, inner and outer integrities, compression, and histological diagnostic value in 3 sets (6 per set) of biopsies for each experimental bioptome. The bioptomes were tested in vivo in 3 healthy minipigs per bioptome. The clinical feasibility was evaluated by procedural and cutting success as well as histological diagnostic value.

Results

The bioptome with the 'grind-grind' design achieved similar values to control in compression (p = 0.822), inner (p = 0.628), and outer (p = 0.507), integrities ex vivo. It showed a better performance in the in vivo real-time MRI setting demonstrating a higher cutting success (91.7%) than the 'grind-anvil' (86.2%) design. In both ex vivo and in vivo evaluations, the 'grind-grind' design displayed sufficient diagnostic value (83% and 95%). The 'grind-anvil' design showed adequate diagnostic value both ex vivo and in vivo (78% and 87.5%) but was not comparable to control according to the three-dimensional (3D) analysis.

Conclusion

A novel MR-compatible bioptome was identified as plausible in a clinical setting. Additionally, SRµCT and subsequent 3D structural analysis could be valuable in the label-free investigation of myocardial tissue at a micrometer level.

Relevance statement

Implementation of MR-guided biopsy can improve animal studies on structural myocardial changes at any point in an experimental setup. With further improvements in guiding catheters, MR-guided biopsy, using the new bioptome, has a potential to increase quality and diagnostic accuracy in patients both with structural and inflammatory cardiomyopathies.

Key points

• Novel MR-compatible bioptomes show promise for a clinical application. • SRµCT enabled detailed analysis of endomyocardial biopsies. • The bioptomes showed adequate in vivo performance without major complications.

SUBMITTER: Svetlove A 

PROVIDER: S-EPMC10695907 | biostudies-literature | 2023 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Evaluation of MR-safe bioptomes for MR-guided endomyocardial biopsy in minipigs: a potential radiation-free clinical approach.

Svetlove Angelika A   Ritter Christian O CO   Dullin Christian C   Schmid Michael M   Schauer Senta S   Uihlein Johannes J   Uecker Martin M   Mietsch Matthias M   Stadelmann Christine C   Lotz Joachim J   Unterberg-Buchwald Christina C  

European radiology experimental 20231205 1


<h4>Background</h4>Diagnostic accuracy of endomyocardial biopsy could improve if clinically safe magnetic resonance (MR)-compatible bioptomes were available. We explored two novel MR-compatible cardiac bioptomes for performance, safety, and clinical viability, employing in vivo minipig trials and phase-contrast synchrotron radiation computed microtomography (SRµCT).<h4>Methods</h4>Analysis of ex vivo obtained pig endomyocardial biopsies was performed using phase-contrast SRµCT and conventional t  ...[more]

Similar Datasets

| S-EPMC5395773 | biostudies-literature
| S-EPMC3504318 | biostudies-other
| S-EPMC9075555 | biostudies-literature
| S-EPMC5908263 | biostudies-literature
| S-EPMC8603053 | biostudies-literature
| S-EPMC10721410 | biostudies-literature
| S-EPMC6153978 | biostudies-literature
| S-EPMC5791188 | biostudies-literature
| S-EPMC4572575 | biostudies-literature
| S-EPMC6059120 | biostudies-literature