An Adjustable Magnetic Levator Prosthesis for Customizable Eyelid Reanimation in Severe Blepharoptosis II: Randomized Evaluation of Angular Translation.


ABSTRACT:

Purpose

Examine the effect of force modulation via angular translation of a static magnetic field for customizable treatment of severe blepharoptosis.

Methods

Prototype adjustable-force magnetic levator prostheses (aMLP) consisted of a spectacle-mounted magnet in rotatable housing and small eyelid-attached magnets embedded in a biocompatible polymer. Interpalpebral fissure (IPF) of 17 participants with severe blepharoptosis was continuously measured for one minute at five spectacle magnet angles, with order randomized and participant and data analyst masked. The hypothesis that angular position affected opening IPF (o-IPF), minimum blink IPF (m-IPF), and comfort ratings (1-10) was tested.

Results

The aMLP improved o-IPF from 4.5 mm without the device to 6.2 mm on the