Ontology highlight
ABSTRACT: Objectives
The aim of this study was to evaluate the performance of a new double-antigen sandwich test that is based on the light-initiated chemiluminescent assay (LiCA®) for detecting anti-hepatitis C virus antibodies (anti-HCV) in comparison to Architect®.Methods
Analytical characteristics and diagnostic performance were tested using seroconversion panels and large pools of clinical samples. Positive results were validated by the strip immunoblot assay (RIBA) and HCV RNA.Results
Repeatability and within-lab imprecision of LiCA® anti-HCV were 1.31%-3.27%. The C5-C95 interval was -5.44%-5.03% away from C50. LiCA® detected seroconversion in an average of 28.9 days and showed a mean of 3.7 (<