ABSTRACT: Purpose
To evaluate the laser power stability of the SubCyclo probe for micropulse transscleral cyclophotocoagulation after repeated use.Materials and methods
This experimental study involved 6 new probes. Each probe was connected to the SubCyclo mode (2,000 mW power, 31.3% duty cycle, and 100 seconds duration) of the Vitra 810 laser delivery system (Quantel Medical, France). Laser power measurements were taken using a calibrated laser power meter (Nova, Ophir Optronics Solutions, Israel) every 10 seconds from 10 to 90 seconds during each of the 40 cycles. Intra-rater reliability was assessed using intraclass correlation (ICC). A linear mixed model for repeated measures and pairwise comparisons with Bonferroni adjustment were used for the analysis.Results
The mean (