Simulated ventilation of two patients with a single ventilator in a pandemic setting.


ABSTRACT: Simultaneous ventilation of two patients may lead to hypoventilation in one patient and hyperinflation in the other patient. In a simulation of ventilation in two patients using artificial lungs, we voluntarily directed gas flow to one patient by using three-dimensional-printed Y-adapters and stenosis adapters during volume- and pressure-controlled ventilation in the first set up. We continuously modified the model using a special one-way valve on the flow-limited side and measured it in pressure-controlled ventilation with the flow sensor of the ventilator adjusted on both sides in a second and third setup. In the first setup, volume- or pressure-controlled ventilation resulted in comparable minute volumes in both lungs, even when one side was obstructed to 3 mm. In the second setup, with

SUBMITTER: Schepat P 

PROVIDER: S-EPMC10715326 | biostudies-literature | 2024 Apr-Jun

REPOSITORIES: biostudies-literature