ABSTRACT: Background
The quick and easy nature of point-of-care (POC) testing devices allows regular monitoring of serum lipid levels to increase efficiency. The purpose of this study was to assess a POC lipid analyzer, Barozen Lipid Plus (MICO Biomed Co., Ltd.), which uses capillary blood to measure total cholesterol (TC), triglycerides (TGs), and high-density lipoprotein cholesterol (HDL-C).Methods
Capillary and venous blood samples were collected from 110 participants at a single center in Korea between June 10 and June 26, 2021. TC, TG, and HDL-C measurements using Barozen Lipid Plus were compared with measurements using our reference device, the Roche-Hitachi Cobas 8000 c702 (Hitachi High-Technologies Corporation). This study followed the guidelines of the Clinical and Laborator
SUBMITTER: Yu S
PROVIDER: S-EPMC10718599 | biostudies-literature | 2023 Nov