Unknown

Dataset Information

Evaluating temporal bone column density for optimized bone conduction implant placement.


ABSTRACT:

Introduction

An optimal placement of bone conduction implants can provide more efficient mechanical transmission to the cochlea if placed in regions with greater bone column density. The aim of this study was to test this hypothesis and to determine the clinical potential of preoperative bone column density assessment for optimal implant placement.

Methods

Five complete cadaver heads were scanned with quantitative computed tomography imaging to create topographic maps of bone density based on the column density index (CODI). Laser Doppler vibrometry was used to measure cochlear promontory acceleration under bone conduction stimulation in different locations on the temporal bone, using a bone-anchored hearing aid transducer at frequencies ranging from 355 Hz to 10 kHz.

Res

SUBMITTER: Talon E 

PROVIDER: S-EPMC10720247 | biostudies-literature | 2023

REPOSITORIES: biostudies-literature