Ontology highlight
ABSTRACT: Background
T2-weighted Single Shot Fast Spin Echo (SSFSE) scans at 3 Tesla (3T) are increasingly used to image fetal pathology due to their excellent tissue contrast resolution and signal-to-noise ratio (SNR). Temperature changes that may occur in response to radio frequency (RF) pulses used for these sequences at 3T have not been studied in human fetal brains. To evaluate the safety of T2-weighted SSFSE for fetal brains at 3T, magnetic resonance (MR) thermometry was used to measure relative temperature changes in a typical clinical fetal brain MR exam.Methods
Relative temperature was estimated using sets of gradient recalled echo (GRE) images acquired before and after T2-weighted SSFSE images which lasted 27.47±8.19 minutes. Thirty-one fetuses with cardiac abnormalities, a
SUBMITTER: Ellison J
PROVIDER: S-EPMC10722078 | biostudies-literature | 2023 Dec
REPOSITORIES: biostudies-literature