Evaluation of the BioFire® FilmArray® Pneumonia plus Panel for Detecting Bacterial Etiological Agents of Lower Respiratory Tract Infections in an Oncologic Hospital. Comparison with Conventional Culture Method.
Ontology highlight
ABSTRACT: Conventional methods used to determine pneumonia pathogens are characterized by low sensitivity and long turnaround times. Introducing new tests with better parameters in patients at higher risk of infections is highly anticipated. The results of the conventional quantitative culture method (CM) in determining the bacterial etiology of pneumonia were compared with the results of the Pneumonia plus Panel test (PNP; BioFire® Diagnostics, USA) in 79 samples of bronchoalveolar lavage (BAL). Materials were collected from 79 patients with suspected pneumonia treated in an oncologic hospital due to solid tumors. Only 16/79 BAL samples (20.3%) were true positive (TP) for bacterial etiology in CM vs. 27/79 samples (34.2%) true positive in the PNP test. The total agreement between methods of
SUBMITTER: Szymankiewicz MT
PROVIDER: S-EPMC10725156 | biostudies-literature | 2023 Dec
REPOSITORIES: biostudies-literature