ABSTRACT: Purpose
To evaluate the resultant thickness of the WaveLight FS200 (Alcon Vision, LLC) created LASIK flap compared to intended thickness at the surgical visit, using the WaveLight EX500 (Alcon Vision, LLC) pachymeter and the resultant flap diameter compared to intended diameter.Methods
This single arm, prospective, single surgeon study assessed the accuracy of the intended flap thickness and diameter, after successful bilateral LASIK surgery. The WaveLight FS200 femtosecond laser was used to create all flaps with an intended thickness of 120 μm. Flap thickness was calculated by subtracting the stromal bed thickness after flap lift from the preoperative corneal thickness using the WaveLight EX500 on-board optical pachymeter. Flap diameter was determined using digital analysi