Quantitative multi-energy micro-CT: A simulation and phantom study for simultaneous imaging of four different contrast materials using an energy integrating detector.
Ontology highlight
ABSTRACT: Emerging from the development of single-energy Computed Tomography (CT) and Dual-Energy Computed Tomography, Multi-Energy Computed Tomography (MECT) is a promising tool allowing advanced material and tissue decomposition and thereby enabling the use of multiple contrast materials in preclinical research. The scope of this work was to evaluate whether a usual preclinical micro-CT system is applicable for the decomposition of different materials using MECT together with a matrix-inversion method and how different changes of the measurement-environment affect the results. A matrix-inversion based algorithm to differentiate up to five materials (iodine, iron, barium, gadolinium, residual material) by applying four different acceleration voltages/energy levels was established. We carried out si
SUBMITTER: Kronfeld A
PROVIDER: S-EPMC10750148 | biostudies-literature | 2024 Jan
REPOSITORIES: biostudies-literature
ACCESS DATA