Ontology highlight
ABSTRACT: Aims
The first-generation radiofrequency HotBalloon (RHB) is a size-adjustable single-shot device used in atrial fibrillation. The energy output is determined by its central temperature and not by its balloon surface temperature (BST), thus limiting its efficacy and safety. Therefore, a second-generation RHB was developed to monitor BST and enable BST-controlled ablation. This animal study aims to evaluate the accuracy of a newly developed BST-monitoring system and validate the optimal BST for ablation.Methods and results
In Protocol 1, thermocapsules were attached to the superior vena cava (SVC) epicardium. The accuracy of BST monitoring was examined during SVC isolation. In Protocol 2, the efficacy and safety of different BST-controlled ablations were examined. In the acu
SUBMITTER: Shinoda Y
PROVIDER: S-EPMC10751809 | biostudies-literature | 2023 Dec
REPOSITORIES: biostudies-literature