Oscillometric measurement of the ankle-brachial index and the estimated carotid-femoral pulse wave velocity improves the sensitivity of an automated device in screening peripheral artery disease.


ABSTRACT:

Background and aims

To overcome the time and personnel constraints of the Doppler method, automated, four-limb blood pressure monitors were recently developed. Their additional functions, such as measuring the estimated carotid-femoral pulse wave velocity (ecfPWV), have been, thus far, less studied. We aimed to compare the sensitivity and specificity of different ankle-brachial index (ABI), toe-brachial index (TBI), and ecfPWV measurement methodologies to evaluate their contribution to peripheral artery disease (PAD) screening.

Methods

Among 230 patients (mean age 64 ± 14 years), ABI measurements were performed using a Doppler device and a manual sphygmomanometer. The Doppler ABI was calculated by taking the higher, while the modified Doppler ABI by taking the lower systolic