Ontology highlight
ABSTRACT: Objective
To systematically evaluate the ability of the CINA® LVO software to detect large vessel occlusions eligible for mechanical thrombectomy on CTA using conventional neuroradiological assessment as gold standard.Methods
Retrospectively, two hundred consecutive patients referred for a brain CTA and two hundred patients that had been subject for endovascular thrombectomy, with an accessible preceding CTA, were assessed for large vessel occlusions (LVO) using the CINA® LVO software. The patients were sub-grouped by occlusion site. The original radiology report was used as ground truth and cases with disagreement were reassessed. Two-by-two tables were created and measures for LVO detection were calculated.Results
A total of four-hundred patients were included; 22
SUBMITTER: Mellander H
PROVIDER: S-EPMC10764253 | biostudies-literature | 2024 Jun