Ontology highlight
ABSTRACT: Background
Vitamin D (vit-D) deficiency is highly prevalent in the Korean population, highlighting the need for accurate measurements. In this study, the interferences by endogenous and cross-reactive substances were compared between routine vit-D immunoassays and mass spectrometry (MS) methods.Methods
Two MS methods and 4 immunoassays from different manufacturers (Abbott, Beckman Coulter, Roche, Siemens) were compared. Residual samples that were icteric, lipemic, hemolyzed, high in rheumatoid factor, from myeloma patients, or patients undergoing hemodialysis were collected. Also, 4 levels of National Institute of Standards and Technology (NIST) Standard Reference Material 972a, and 12 samples serially spiked with 3-epi-25-OH-D3 were prepared.Results
Sign
SUBMITTER: Lee JH
PROVIDER: S-EPMC10770599 | biostudies-literature | 2024 Jan