Mixed Reality Surgical Navigation System; Positional Accuracy Based on Food and Drug Administration Standard.


ABSTRACT:

Background

Computer assisted surgical navigation systems are designed to improve outcomes by providing clinicians with procedural guidance information. The use of new technologies, such as mixed reality, offers the potential for more intuitive, efficient, and accurate procedural guidance. The goal of this study is to assess the positional accuracy and consistency of a clinical mixed reality system that utilizes commercially available wireless head-mounted displays (HMDs), custom software, and localization instruments.

Methods

Independent teams using the second-generation Microsoft HoloLens© hardware, Medivis SurgicalAR© software, and localization instruments, tested the accuracy of the combined system at different institutions, times, and locations. The

SUBMITTER: Morley CT 

PROVIDER: S-EPMC10773158 | biostudies-literature | 2024 Feb

REPOSITORIES: biostudies-literature