Adsorptive stripping voltammetric sensor based on Cd zeolitic imidazole framework-67 for electrochemical detection of sarin simulant.
Ontology highlight
ABSTRACT: A selective and reliable modified glassy carbon sensor, based on a 1.0% Cd zeolitic imidazole framework-67 modified glassy carbon sensor (GCS2), has been developed for ultrasensitive detection of dimethyl methyl phosphonate (DMMP) in human biological fluid. The synthesis of porous nanoparticles of Cd zeolitic imidazole framework-67 (Cd ZIF-67) was carried out via the hydrothermal method. The resulting Cd ZIF-67 powder emerges with good crystallinity, a rhombic dodecahedral morphology with particle size in the range 300 ~ 500 nm, and a specific surface area of 1780 m2·g-1. Furthermore, the fabricated sensor exhibited superior performance in the detection of DMMP with two linearity ranges of 0.02-2.0 nM and 2.0-9.0 nM and a limit of detection (LOD) of 0.06 pM
SUBMITTER: Elfiky M
PROVIDER: S-EPMC10774163 | biostudies-literature | 2024 Jan
REPOSITORIES: biostudies-literature