ABSTRACT: Background
Using a fluid-filled wire with a pressure sensor outside the patient compared to a conventional pressure wire may avoid the systematic error introduced by the hydrostatic pressure within the coronary circulation.Aims
To assess the safety and effectiveness of the novel fluid-filled wire, Wirecath (Cavis Technologies, Uppsala, Sweden), as well as its ability to avoid the hydrostatic pressure error.Methods and results
The Wirecath pressure wire was used in 45 eligible patients who underwent invasive coronary angiography and had a clinical indication for invasive coronary pressure measurement at Sahlgrenska University Hospital, Gothenburg, Sweden. In 29 patients, a simultaneous measurement was performed with a conventional coronary pressure wire (PressureWire