ABSTRACT: Introduction
The aim of this study was to investigate the influence of ocular biometry parameters on the predictive accuracy of 10 intraocular lens (IOL) power formulas in patients with high myopia (HM).Methods
We analyzed 202 eyes of 202 patients. The ocular biometry was determined preoperatively using an IOLMaster 700. The associations between the biometry parameters and the prediction error (PE) 1 month postoperatively were assessed. HM was defined as an axial length exceeding 26.50 mm.Results
In patients with HM (n = 108), the K6, Emmetropia Verifying Optical (EVO), Olsen, and Barrett Universal II (BUII) formulas had the lowest absolute PEs among the 10 formulas. The ocular biometry parameters were not associated with the PE of K6, EVO, Olsen, or BUII. A longer
SUBMITTER: Miao A
PROVIDER: S-EPMC10776536 | biostudies-literature | 2024 Jan