Ontology highlight
ABSTRACT: Background
Recently, a ring-type cuffless blood pressure (BP) measuring device has been developed. This study was a prospective, single arm, first-in-human pivotal trial to evaluate accuracy of BP measurement by the new device.Methods
The ring-type smart wearable monitoring device measures photoplethysmography signals from the proximal phalanx and transmits the data wirelessly to a connected smartphone. For the BP comparison, a cuff was worn on the arm to check the reference BP by auscultatory method, while the test device was worn on the finger of the opposite arm to measure BP simultaneously. Measurements were repeated for up to three sets each on the left and right arms. The primary outcome measure was mean difference and standard deviation of BP differences between the
SUBMITTER: Kim J
PROVIDER: S-EPMC10789523 | biostudies-literature | 2024 Jan
REPOSITORIES: biostudies-literature