Finite element analysis and clinical application of 3D-printed Ti alloy implant for the reconstruction of mandibular defects.


ABSTRACT:

Background

The reconstruction of segmental defect of the mandible has always been a challenge. The customized reconstruction plate without a bone graft is also considered a transitional means of rehabilitation and reconstruction in some cases.

Methods

This study evaluated the biomechanical behaviors of customized plates with different structural designs comparing with commercial plates using the finite element method in reconstrution of the lateral mandible defect.

Results

Simulations revealed the stress state in the plate bodies, bone tissues and screws were associated with the width, height, thickness of the plates as well as the distribution of screws. In all of the groups, the system of 16 mm-high, 2.8 mm-thick customized reconstruction plate with 10 screws was co

SUBMITTER: Xue R 

PROVIDER: S-EPMC10792868 | biostudies-literature | 2024 Jan