Real-World Use of Control-IQ Technology Is Associated with a Lower Rate of Severe Hypoglycemia and Diabetic Ketoacidosis Than Historical Data: Results of the Control-IQ Observational (CLIO) Prospective Study.
Ontology highlight
ABSTRACT: Objective: Severe hypoglycemia (SH) and diabetic ketoacidosis (DKA) remain significant risks with intensive insulin therapy. While these adverse event (AE) rates are generally very low in advanced hybrid closed-loop (AHCL) clinical studies, prospectively collected real-world AE rates are lacking. Research Design and Methods: The Control-IQ Observational (CLIO) study was a single-arm, prospective, longitudinal, postmarket surveillance study of individuals with type 1 diabetes (T1D) age 6 years and older who began the use of t:slim X2 insulin pump with Control-IQ technology in the real-world outpatient setting. AEs were reported monthly over 12 months and were compared to historical data from the T1D Exchange. Patient-reported outcomes were assessed quarterly. All