ABSTRACT: Purpose
To investigate 2-year post-operative hearing performance, safety, and patient-reported outcomes of hearing-impaired adults treated with the Osia® 2 System, an active osseointegrated bone-conduction hearing implant that uses piezoelectric technology.Methods
A prospective, multicenter, open-label, single-arm, within-subject clinical study conducted at three tertiary referral clinical centers located in Melbourne, Sydney and Hong Kong. Twenty adult recipients of the Osia 2 System were enrolled and followed up between 12 and 24 months post-implantation: 17 with mixed or conductive hearing loss and 3 with single-sided sensorineural deafness. Safety data, audiological thresholds, speech recognition thresholds in noise, and patient-reported outcomes were collect