Numerical investigation reveals challenges in measuring the contrast recovery coefficients in PET.
Ontology highlight
ABSTRACT: Objective.Contrast recovery coefficient (CRC) is essential for image quality (IQ) assessment in positron emission tomography (PET), typically measured according to the National Electrical Manufacturers Association (NEMA) NU 2 standard. This study quantifies systematic uncertainties of the CRC measurement by a numerical investigation of the effects from scanner-independent parameters like voxel size, region-of-interest (ROI) misplacement, and sphere position on the underlying image grid.Approach.CRC measurements with 2D and 3D ROIs were performed on computer-generated images of a NEMA IQ-like phantom, using voxel sizes of 1-4 mm for sphere diameters of 5-40 mm-first in absence of noise and blurring, then with simulated spatial resolution and image noise with varying noise leve