Unknown

Dataset Information

LC-ESI-HRMS - lipidomics of phospholipids : Characterization of extraction, chromatography and detection parameters.


ABSTRACT: Lipids are a diverse class of molecules involved in many biological functions including cell signaling or cell membrane assembly. Owing to this relevance, LC-MS/MS-based lipidomics emerged as a major field in modern analytical chemistry. Here, we thoroughly characterized the influence of MS and LC settings - of a Q Exactive HF operated in Full MS/data-dependent MS2 TOP N acquisition mode - in order to optimize the semi-quantification of polar lipids. Optimization of MS-source settings improved the signal intensity by factor 3 compared to default settings. Polar lipids were separated on an ACQUITY Premier CSH C18 reversed-phase column (100 × 2.1 mm, 1.7 µm, 130 Å) during an elution window of 28 min, leading to a sufficient number of both data points across the chromatographic pea

SUBMITTER: Rund KM 

PROVIDER: S-EPMC10800306 | biostudies-literature | 2024 Feb

REPOSITORIES: biostudies-literature