Unknown

Dataset Information

Cat alpha-1-acid glycoprotein enzyme-linked immunosorbent assay: performance characteristics and reference intervals.


ABSTRACT:

Objectives

The aim of this study was to evaluate performance of a feline-specific ELISA for the measurement of alpha-1-acid glycoprotein (AGP) and to establish a reference interval (RI).

Methods

Surplus serum samples with low (~200 µg/ml), medium (~450 µg/ml) and high (~745 and 930 µg/ml) AGP concentrations were used to assess the intra- and inter-assay coefficients of variation (CVs). The quality goal for bioanalytical method validation was a CV of <20%. Linearity was assessed by serial dilution of a sample with a high AGP concentration. Spike recovery was evaluated by mixing samples with low, medium and high AGP concentrations at different ratios. To establish the RI, residual serum samples from 51 healthy adult cats that were presented for health examinations or blood don

SUBMITTER: Thalmeier S 

PROVIDER: S-EPMC10811989 | biostudies-literature | 2023 May

REPOSITORIES: biostudies-literature