Valve Hemodynamics by Valve Size and 1-Year Survival Following Implantation of the Portico Valve in the Multicenter CONFIDENCE Registry.


ABSTRACT:

Background

The CONtrolled delivery For ImproveD outcomEs with cliNiCal Evidence registry was initiated to characterize the clinical safety and device performance from experienced transcatheter aortic valve implantation (TAVI) centers in Europe and Australia that use the Portico valve to treat patients with severe aortic stenosis. We herein report for the first time the valve performance at 30-day across all implanted valve sizes and the 1-year survival from this registry.

Methods

This was a prospective, multicenter, single-arm observational clinical investigation of patients clinically indicated for implantation of a Portico valve in experienced TAVI centers. Patients were treated with a commercially available valve (size 23, 25, 27, or 29 mm) using either the first-generati