First experimental demonstration of real-time neutron capture discrimination in helium and carbon ion therapy.


ABSTRACT: This work provides the first experimental proof of an increased neutron capture photon signal following the introduction of boron to a PMMA phantom during helium and carbon ion therapies in Neutron Capture Enhanced Particle Therapy (NCEPT). NCEPT leverages [Formula: see text]B neutron capture, leading to the emission of detectable 478 keV photons. Experiments were performed at the Heavy Ion Medical Accelerator in Chiba, Japan, with two Poly(methyl methacrylate) (PMMA) targets, one bearing a boron insert. The BeNEdiCTE gamma-ray detector measured an increase in the 478 keV signal of 45 ± 7% and 26 ± 2% for carbon and helium ion irradiation, respectively. Our Geant4 Monte Carlo simulation model, developed to investigate photon origins, found less than 30% of detected photons originated from

SUBMITTER: Kielly M 

PROVIDER: S-EPMC10831067 | biostudies-literature | 2024 Jan

REPOSITORIES: biostudies-literature