ABSTRACT: Introduction
This study evaluated the performance of the Lumipulse plasma beta-amyloid (Aβ) 42/40 and pTau181 compared to other assays to detect an abnormal amyloid-positron emission tomography (PET).Methods
Plasma samples from cognitively unimpaired (N = 179) and MCI/AD dementia (N = 36) individuals were retrospectively evaluated. Plasma Aβ42/40 and pTau181 were measured using the Lumipulse and Simoa immunoassays. An immunoprecipitation mass spectrometry (IP-MS) assay for plasma Aβ42/40 was also evaluated. Amyloid-PET status was the outcome measure.Results
Lumipulse and IP-MS Aβ42/40 exhibited the highest diagnostic accuracy for detecting an abnormal amyloid-PET (areas under the curve [AUCs] of 0.81 and 0.84, respectively). The Lumipulse and Simoa pTa