ABSTRACT: Purpose
To determine the feasibility of replacing the mid treatment cone beam computed tomography (MT CBCT) image with Intrafraction Imaging (IFI) acquired concurrently during dose delivery in lung Stereotactic Ablative Body Radiation therapy (SABR) patients, and thus improve treatment efficiency.Methods and materials
A review of departmental imaging data was performed on ten lung SABR patients treated with dual arc volumetric modulated arc therapy (VMAT) on an Elekta Versa HD linear accelerator with XVI imaging software.IFI data was extracted and a database of the translational (TX, TY, TZ) and the rotational (RX, RY, RZ) position errors was created for retrospective comparison, with the values of the MT CBC