ABSTRACT: Background
Continuous renal replacement therapy (CRRT) support is crucial for critically ill patients and it is underexplored in specific situations. Experimental CRRT offers a means to gain insights into these scenarios, but the prohibitive cost of CRRT machines limits their accessibility. This study aimed to develop and validate a low-cost and precise dialysate controller for experimental CRRT.Results
Our results demonstrate a commendable level of precision in affluent flow control, with a robust correlation (R2 = 0.99) for continuous flow and a strong correlation (R2 = 0.95) for intermittent flow. Additionally, we observed acceptable agreement with a bias = 3.4 mL (upper limit 95% = 43.9 mL and lower limit 95% = - 37 mL) for continuous flow and bias