ABSTRACT: Introduction
Sufficiently high analytical quality of blood glucose monitoring systems (BGMS) is a prerequisite for efficient diabetes therapy. In this study we assessed system accuracy, measurement repeatability, intermediate measurement precision, user performance, and the influence of hematocrit on two CE-marked blood glucose monitoring systems. For one BGMS, measurement accuracy using venous samples was additionally investigated.Methods
Study procedures were based on the International Organization of Standardization (ISO) 15197:2013/EN ISO 15197:2015 ("ISO 15197"). User performance included data from 100 subjects who used one test strip lot, whereas for all other analyses three different reagent system lots were used. For system accuracy assessment, 100 capillary samples