The retropharyngeal reduction plate for atlantoaxial dislocation: a finite element analysis.


ABSTRACT: Objective: To investigate the biomechanical properties of the retropharyngeal reduction plate by comparing the traditional posterior pedicle screw-rod fixation by finite element analysis. Methods: Two three-dimensional finite element digital models of the retropharyngeal reduction plate and posterior pedicle screw-rod fixation were constructed and validated based on the DICOM (Digital Imaging and Communications in Medicine) data from C1 to C4. The biomechanical finite element analysis values of two internal fixations were measured and calculated under different conditions, including flexion, extension, bending, and rotation. Results: In addition to the backward extension, there was no significant difference in the maximum von Mises stress between the retropharyngeal re