Evaluation and experience from routine use of chemiluminescence assays for serological screening of blood and plasma donations on the Alinity s system and the Alinity i system, two new fully-automated immunoassay systems in Poland.
Ontology highlight
ABSTRACT: In Poland, independent evaluations under the auspices of the Institute of Hematology and Transfusion Medicine (IHTM) are mandated for any new device, assay, systems for screening samples from whole blood and plasma donors prior to implementation by Blood Transfusion Center (BTC). In last 5 years, two new systems were introduced to the market by Abbott GmbH, namely the Alinity s and the Alinity i. The evaluations performed for these two systems included the assessment of sensitivity, specificity and precision for each of the four mandatory serological screening markers in Poland: Hepatitis B Surface Antigen (HBsAg), Hepatitis C virus antibodies (Anti-HCV), HIV antibodies (anti-HIV) and Syphilis antibodies (anti-Treponema pallidum, anti-TP). Sensitivity was assessed by testing serocon
SUBMITTER: Kopacz A
PROVIDER: S-EPMC10847982 | biostudies-literature | 2024 Mar
REPOSITORIES: biostudies-literature