Assessment of novel POCT to evaluate liver function.


ABSTRACT:

Objectives

Point of care testing (POCT) offers the possibility of near bedside patient testing with a reduction of the turn-around time of analysis. The aim of our study was to determine the analytical performances and usability of a recently developed POCT device for the measurement of tests related to liver function. We evaluated the performance of a liver tests panel performed on the LINX EVO® POCT device.

Design and methods

The imprecision was determined with the Bio-Rad Liquichek Unassayed Chemistry Control. Method comparison was performed with a Cobas® 8000 analyzer. Samples from twenty healthy volunteers were used to verify the reference intervals. Furthermore, practicality was assessed by the healthcare staff handling the POCT device through a dedicated questionnaire