Ontology highlight
ABSTRACT: Background
There is no standard method for assembling the femoral head onto the femoral stem during total hip arthroplasty (THA). This study aimed to measure and record dynamic 3-dimensional (3D) THA head-neck assembly loads from residents, fellows, and attending surgeons, for metal and ceramic femoral heads.Methods
An instrumented apparatus measured dynamic 3D forces applied through the femoral stem taper in vitro for 31 surgeons (11 attendings, 14 residents, 6 fellows) using their preferred technique (ie, number of hits or mallet strikes). Outcome variables included peak axial force, peak resultant force, impulse of the resultant force, loading rate of the resultant force, and off-axis angle. They were compared between femoral head material, surgeon experience level, and
SUBMITTER: Godoy M
PROVIDER: S-EPMC10850914 | biostudies-literature | 2024 Feb