Ontology highlight
ABSTRACT: Background
To describe the development and investigate the accuracy of a novel smartphone-based Contrast Sensitivity (CS) application, the K-CS test.Methods
A total of 67 visually impaired and 50 normal participants were examined monocularly using the novel digital K-CS test and the Pelli-Robson (PR) chart. The K-CS test examines letter contrast sensitivity in logarithmic units, using eight levels of contrast from logCS = ~0,1 to logCS = ~2,1 at two spatial frequencies of 1.5 and 3 cycles per degree (cpd). The K-CS test was compared to the gold standard, PR test and intra-session test repeatability was also examined.Results
The K-CS test in normally sighted was found to agree well with the PR, providing comparable mean scores in logCS (±SD) (K-CS = 1.908 ± 0.06 vers
SUBMITTER: Karampatakis V
PROVIDER: S-EPMC10852338 | biostudies-literature | 2024
REPOSITORIES: biostudies-literature