Evaluation of contrast sensitivity in visually impaired individuals using K-CS test. A novel smartphone-based contrast sensitivity test-Design and validation.


ABSTRACT:

Background

To describe the development and investigate the accuracy of a novel smartphone-based Contrast Sensitivity (CS) application, the K-CS test.

Methods

A total of 67 visually impaired and 50 normal participants were examined monocularly using the novel digital K-CS test and the Pelli-Robson (PR) chart. The K-CS test examines letter contrast sensitivity in logarithmic units, using eight levels of contrast from logCS = ~0,1 to logCS = ~2,1 at two spatial frequencies of 1.5 and 3 cycles per degree (cpd). The K-CS test was compared to the gold standard, PR test and intra-session test repeatability was also examined.

Results

The K-CS test in normally sighted was found to agree well with the PR, providing comparable mean scores in logCS (±SD) (K-CS = 1.908 ± 0.06 versus PR = 1.93 ± 0.05) at 1.5 cpd and mean (± SD) logCS at 3 cpd (K-CS = 1.83 ± 0.13 versus PR = 1.86 ± 0.07). The mean best corrected visual acuity of visually impaired participants was 0.67 LogMAR (SD = 0.21) and the K-CS was also found to agree well with the Pelli-Robson in this group, with an equivalent mean (±SD) logCS at 1.5 cpd: (K-CS = 1.19 ± 0.27, PR = 1.15 ± 0.31), 3 cpd: K-CS = 1.01 ± 0.33, PR = 0.94 ± 0.34. Regarding the intra-session test repeatability, both the K-CS test and the PR test showed good repeatability in terms of the 95% limits of agreement (LoA): K-CS = ±0.112 at 1.5 cpd and ±0.133 at 3 cpd, PR = ±0.143 at 1.5 cpd and ±0.183 in 3 cpd in visually impaired individuals.

Conclusion

The K-CS test provides a quick assessment of the CS both in normally sighted and visually impaired individuals. The K-CS could serve as an alternative tool to assess contrast sensitivity function using a smartphone and provides results that agree well with the commonly used PR test.

SUBMITTER: Karampatakis V 

PROVIDER: S-EPMC10852338 | biostudies-literature | 2024

REPOSITORIES: biostudies-literature

altmetric image

Publications

Evaluation of contrast sensitivity in visually impaired individuals using K-CS test. A novel smartphone-based contrast sensitivity test-Design and validation.

Karampatakis Vasileios V   P Papadopoulou Eleni E   Almpanidou Stavroula S   Karamitopoulos Leonidas L   Almaliotis Diamantis D  

PloS one 20240208 2


<h4>Background</h4>To describe the development and investigate the accuracy of a novel smartphone-based Contrast Sensitivity (CS) application, the K-CS test.<h4>Methods</h4>A total of 67 visually impaired and 50 normal participants were examined monocularly using the novel digital K-CS test and the Pelli-Robson (PR) chart. The K-CS test examines letter contrast sensitivity in logarithmic units, using eight levels of contrast from logCS = ~0,1 to logCS = ~2,1 at two spatial frequencies of 1.5 and  ...[more]

Similar Datasets

| S-EPMC6743644 | biostudies-literature
| S-EPMC10037403 | biostudies-literature
| S-EPMC10849460 | biostudies-literature
| S-EPMC6137691 | biostudies-literature
| S-EPMC6844543 | biostudies-literature
| S-EPMC5696488 | biostudies-literature
| S-EPMC11684699 | biostudies-literature
| S-EPMC7396172 | biostudies-literature
| S-EPMC9044340 | biostudies-literature
| S-EPMC5794170 | biostudies-literature