Evaluation of immature granulocyte parameters in myeloid neoplasms assayed by Sysmex XN hematology analyzer.


ABSTRACT: Immature granulocytes (IGs) have significance for the diagnosis of myeloid neoplasms (MNs). The current study aims to use a hematology analyzer to evaluate the accuracy of IG parameters in MNs. Blood specimens from 388 patients with MN, 524 with non-hematological neoplasms (non-HNs), including 109 patients with inflammation and 68 undergoing G-CSF administration, and 500 healthy control subjects were analyzed. IG parameters was assayed by Sysmex XN-9000 (XN) and compared with manual assessments. A high level of agreement between IG% derived from XN and manual measurements for MN patients (r = 0.828, p < 0.0001) was revealed but only a moderate correlation for acute myeloid leukemia patients (AML; r = 0.597; p < 0.0001). Bland-Altman bias analysis was conducted, and the results showed that differences in IG% from XN and manual analysis for MN patients were considered clinically insignificant. ROC analysis demonstrated a good performance of IG# (AUC = 0.842) and IG% (AUC = 0.885) assessed by XN for MN patients with cut-off values of 0.200 × 109/L and 1.95%, respectively. IG parameters from Sysmex XN analyzer are helpful for screening of MNs even though granulocyte morphological abnormalities may interfere with IG parameter accuracy.

SUBMITTER: Lu Q 

PROVIDER: S-EPMC10869398 | biostudies-literature | 2022 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Evaluation of immature granulocyte parameters in myeloid neoplasms assayed by Sysmex XN hematology analyzer.

Lu Qifeng Q   Li Ying Y   Li Tian T   Hou Tingting T   Zhao Yajuan Y   Feng Shu S   Yang Xixian X   Zhu Mengyu M   Shen Yajuan Y  

Journal of hematopathology 20220208 1


Immature granulocytes (IGs) have significance for the diagnosis of myeloid neoplasms (MNs). The current study aims to use a hematology analyzer to evaluate the accuracy of IG parameters in MNs. Blood specimens from 388 patients with MN, 524 with non-hematological neoplasms (non-HNs), including 109 patients with inflammation and 68 undergoing G-CSF administration, and 500 healthy control subjects were analyzed. IG parameters was assayed by Sysmex XN-9000 (XN) and compared with manual assessments.  ...[more]

Similar Datasets

| S-EPMC8362000 | biostudies-literature
| S-EPMC7891428 | biostudies-literature
| S-EPMC8366244 | biostudies-literature
| S-EPMC10267328 | biostudies-literature
| S-EPMC11885685 | biostudies-literature
| S-EPMC2829900 | biostudies-literature
| S-EPMC10812539 | biostudies-literature
| S-EPMC6505781 | biostudies-literature
| S-EPMC5919509 | biostudies-literature
| S-EPMC6543632 | biostudies-literature