Agreement between invasive and noninvasive measurement of tear film breakup time.
Ontology highlight
ABSTRACT: The agreement between fluorescein breakup time (FBUT) estimates and noninvasive breakup time (NIBUT) estimates automatically acquired using two videokeratoscopes was assessed. Small-cone (E300, Medmont) and large-bowl (K5M, OCULUS) Placido-ring-based videokeratoscopes were utilized for automated NIBUT estimation and fluorescein strip method was used for FBUT estimation. 33 subjects completed the study. Clear instructions were given regarding the blink before measurements. Bland-Altman analysis was used to test the agreement between tear film breakup time estimates and repeated measure analysis of variance was used to test the differences between measurement types. In comparison to FBUT, E300 NIBUT estimate was shorter (- 0.6 ± 2.6 s), whereas K5M NIBUT estimate was longer (3.3 ± 2.4 s). Li