Performance Evaluation of Microscanner Plus, an Automated Image-Based Cell Counter, for Counting CD4+ T Lymphocytes in HIV Patients.
Ontology highlight
ABSTRACT: Counting CD4+ T lymphocytes using flow cytometry is a standard method for monitoring patients with HIV infections. Simpler and cheaper alternatives to flow cytometry are in high demand because getting access to flow cytometers is difficult or impossible in resource-limited settings. We evaluated the performance of the Microscanner Plus, a simple and automated image-based cell counter, in determining CD4 counts against a flow cytometer. CD4 count results of the Microscanner Plus and flow cytometer were compared using samples from 47 HIV-infected patients and 87 healthy individuals. All CV% for precision and reproducibility tests were less than 10%. The Microscanner Plus's lowest detectable CD4 count was determined to be 15.27 cells/µL of whole blood samples. The correlation coefficient (R)