Hippocampal single-voxel MR spectroscopy with a long echo time at 3 T using semi-LASER sequence.
Ontology highlight
ABSTRACT: The hippocampus is one of the most challenging brain regions for proton MR spectroscopy (MRS) applications. Moreover, quantification of J-coupled species such as myo-inositol (m-Ins) and glutamate + glutamine (Glx) is affected by the presence of macromolecular background. While long echo time (TE) MRS eliminates the macromolecules, it also decreases the m-Ins and Glx signal and, as a result, these metabolites are studied mainly with short TE. Here, we investigate the feasibility of reproducibly measuring their concentrations at a long TE of 120 ms, using a semi-adiabatic localization by adiabatic selective refocusing (sLASER) sequence, as this sequence was recently recommended as a standard for clinical MRS. Gradient offset-independent adiabatic refocusing pulses were implemented, and an o
SUBMITTER: Gajdosik M
PROVIDER: S-EPMC10874619 | biostudies-literature | 2021 Aug
REPOSITORIES: biostudies-literature