ABSTRACT: Purpose
To compare the effect of capsular bend on the rotational stability between 2 toric intraocular lenses (IOLs).Setting
Eye Hospital of Wenzhou Medical University, Hangzhou, Zhejiang Province, China.Design
Prospective study.Methods
Patients with preexisting astigmatism received AcrySof IQ (SN6AT) or TECNIS (ZCT/ZMT) toric IOL during cataract surgery. CASIA2 was used to record the toric IOL axial orientation and capsular bend index (CBI) at the 1-day, 1-week, 1-month, and 3-month interval postoperatively. The postoperative rotational stability and CBI of both models were compared.Results
A total of 58 eyes from 58 patients were enrolled in this study. The total misalignment of the TECNIS (ZCT/ZMT) group (6.96 ± 5.10 degrees, 7.41 ± 5.19 degrees,