Unknown

Dataset Information

Safety of ventricular arrhythmia radiofrequency ablation with half-normal saline irrigation.


ABSTRACT:

Aims

Failure of radiofrequency (RF) ablation of ventricular arrhythmias is often due to inadequate lesion size. Irrigated RF ablation with half-normal saline (HNS) has the potential to increase lesion size and reduce sodium delivery to the patient if the same volume of RF irrigant were used for normal saline (NS) and HNS but could increase risks related to steam pops and lesion size. This study aims to assess periprocedural complications and acute ablation outcome of ventricular arrhythmias ablation with HNS.

Methods and results

Prospective assessment of outcomes was performed in 1024 endocardial and/or epicardial RF ablation procedures in 935 consecutive patients (median age 64 years, 71.2% men, 73.4% cardiomyopathy, 47.2% sustained ventricular tachycardia). Half-normal sal

SUBMITTER: Hasegawa K 

PROVIDER: S-EPMC10898929 | biostudies-literature | 2024 Feb

REPOSITORIES: biostudies-literature