ABSTRACT: Aim
To evaluate the accuracy of the raytracing method for the calculation of intraocular lens (IOL) power in myopic eyes after small incision extraction of the lenticule (SMILE).Methods
Retrospective study. All patients undergoing surgery for myopic SMILE between May 1, 2020, and December 31, 2020, with Scheimpflug tomography optical biometry were eligible for inclusion. Manifest refraction was performed before and 6 months after refractive surgery. One eye from each patient was included in the final analysis. A theoretical model was invited to predict the accuracy of multiple methods of lens power calculation by comparing the IOL-induced refractive error at the corneal plane (IOL-Dif) and the SMILE-induced change of spherical equivalent (SMILE-Dif) before and after SMILE surgery. The prediction error (PE) was calculated as the difference between SMILE-Dif-IOL-Dif. IOL power calculations were performed using raytracing (Olsen Raytracing, Pentacam AXL, software version 1.22r05, Wetzlar, Germany) and other formulae with historical data (Barrett True-K, Double-K SRK/T, Masket, Modified Masket) and without historical data (Barrett True-K no history, Haigis-L, Hill Potvin Shammas PM, Shammas-PL) for the same IOL power and model. In addition, subgroup analysis was performed in different anterior chamber depths, axial lengths, back-to-front corneal radius ratio, keratometry, lens thickness, and preoperative spherical equivalents.Results
A total of 70 eyes of 70 patients were analyzed. The raytracing method had the smallest mean absolute PE (0.26 ± 0.24 D) and median absolute PE (0.16 D), and also had the largest percentage of eyes within a PE of ± 0.25 D (64.3%), ± 0.50 D (81.4%), ± 0.75 D (95.7%), and ± 1.00 D (100.0%). The raytracing method was significantly better than Double-K SRK/T, Haigis, Haigis-L, and Shammas-PL formulae in postoperative refraction prediction (all p < 0.001), but not better than the following formulae: Barrett True-K (p = 0.314), Barrett True-K no history (p = 0.163), Masket (p = 1.0), Modified Masket (p = 0.806), and Hill Potvin Shammas PM (p = 0.286). Subgroup analysis showed that refractive outcomes exhibited no statistically significant differences in the raytracing method (all p < 0.05).Conclusion
Raytracing was the most accurate method in predicting target refraction and had a good consistency in calculating IOL power for myopic eyes after SMILE.
SUBMITTER: Wei Y
PROVIDER: S-EPMC10898958 | biostudies-literature | 2024 Feb
REPOSITORIES: biostudies-literature
Klinische Monatsblatter fur Augenheilkunde 20230918 2
<h4>Aim</h4>To evaluate the accuracy of the raytracing method for the calculation of intraocular lens (IOL) power in myopic eyes after small incision extraction of the lenticule (SMILE).<h4>Methods</h4>Retrospective study. All patients undergoing surgery for myopic SMILE between May 1, 2020, and December 31, 2020, with Scheimpflug tomography optical biometry were eligible for inclusion. Manifest refraction was performed before and 6 months after refractive surgery. One eye from each patient was ...[more]