Rapid patient-specific FEM meshes from 3D smart-phone based scans.


ABSTRACT: Objective.The objective of this study was to describe and evaluate a smart-phone based method to rapidly generate subject-specific finite element method (FEM) meshes. More accurate FEM meshes should lead to more accurate thoracic electrical impedance tomography (EIT) images.Approach.The method was evaluated on an iPhone®that utilized an app called Heges, to obtain 3D scans (colored, surface triangulations), a custom belt, and custom open-source software developed to produce the subject-specific meshes. The approach was quantitatively validated via mannequin and volunteer tests using an infrared tracker as the gold standard, and qualitatively assessed in a series of tidal-breathing EIT images recorded from 9 subjects.Main results.The subject-specific meshes c