Frequency of Leaks from Conical Centrifuge Tubes.


ABSTRACT:

Introduction

Threaded conical centrifuge tubes are ubiquitous in biological laboratories and are frequently used for the storage/transport of potentially biohazardous samples. However, limited data are available on how frequently and from where these tubes leak. These data are valuable for laboratory biorisk management and to inform future studies on risks arising from the routine use of laboratory consumables.

Methods

The frequency of leaks from threaded conical centrifuge tubes was tested using a Glo Germ solution as a tracer. Conical tubes (15 and 50 mL) from several brands were filled, inverted, and placed on their side on the benchtop. After 1 h, the presence or absence of leaks on the benchtop surface, tube threads, and exterior was recorded.

Results

We observed

SUBMITTER: Wyneken H 