ABSTRACT: The H295R test guideline assay evaluates the effect of test substances on synthesis of 17β-estradiol (E2) and testosterone (T). The objective of this study was to leverage commercial immunoassay technology to develop a more efficient H295R assay to measure E2 and T levels in 384-well format. The resulting Homogenous Time Resolved Fluorescence assay platform (H295R-HTRF) was evaluated against a training set of 36 chemicals derived from the OECD inter-laboratory validation study, EPA guideline 890.1200 aromatase assay, and azole fungicides active in the HT-H295R assay. Quality control performance criteria were met for all conditions except E2 synthesis inhibition where low basal hormone synthesis was observed. Five proficiency chemicals were active for both the E2 and T endpoints, consistent