Data analytics improves the diagnostic accuracy of serum free light chain results for detecting monoclonal gammopathy.


ABSTRACT:

Objectives

To evaluate the real-world performance and reference intervals of the Binding Site Freelite serum free light chain (SFLC) assay (Thermo Fisher Scientific), a global standard for diagnosis, prognostication, and response assessment for monoclonal gammopathies.

Methods

An informatics-based approach was used to retrospectively evaluate concordance between SFLC and the orthogonal Sebia HYDRASYS immunofixation assay results in a large clinical data set consecutively reported between 2010 and 2020.

Results

Among patients with monoclonal-negative results by both SFLC and Sebia HYDRASYS immunofixation assays, 25% (1226/5057) had κ/λ ratios (KLRs) outside the manufacturer-defined and International Myeloma Working Group-cited normal reference interval of 0.26 to 1.65.